Collection Substrate

SEM Stubs or custom liquid petri dishes

Particle Size Range & Collection Efficiency 10nm-10µm>95% @ 4LPM
10nm-10µm >90% @ 6LPM
Additional conditions mid-testing, contact Handix Scientific for more information

Particle Concentration Range

0 to 10^5 #/cc

Condensing Fluid

Water, distilled or higher purity

Sample Flow Rate Range

Liquid Sampling: up to 4 LPM
Solid Sampling: 4-10 LPM
Collection efficiency curve varies with flow rate, contact Handix Scientific for exact information.

Sampled Aerosol Conditions

0-45C, 25-95%RH

Environmental Operating Conditions

5-35 C

Dimensions

Height 9”, Width 8”, Depth 15”

Weight

11 pounds

Specifications subject to change without notice. Aerosol particle collector technology is licensed from NIOSH/CDC
(Zervaki and Kulkarni, Patent Application #US20250130140A1) 

References

Zervaki O., Dionysiou D. D., & Kulkarni P. (2024). Compact, high-flow, water-based, turbulent-mixing, condensation aerosol concentrator for collection of spot samples. Aerosol Science and Technology, 58(8), 889–901. https://doi.org/10.1080/02786826.2024.2361050

Zervaki O., Dionysiou D. D., & Kulkarni P. (2024). A High-Throughput, Turbulent-Mixing, Condensation Aerosol Concentrator for Direct Aerosol Collection as a Liquid Suspension. https://stacks.cdc.gov/view/cdc/208350