| Aerosol Collection | Mixing Condensation Aerosol Collector (MCAC) |
| Sampling Flow Rate | 1–10 LPM |
| MCAC Collection Efficiency | >95% for 8–300 nm particles. Additional conditions mid-testing, contact Handix Scientific for more information. |
| Collection Electrode | Rotating tungsten electrode |
| Measurement | Raman + Spark Emission Spectroscopy |
|
Raman Excitation |
Dual wavelength, 784.7 nm and 785.2 nm |
| Raman Spectral Range | 200–3200 cm⁻¹ |
| Raman Resolution | 4.3 cm⁻¹ |
| Spark Emission Range | 245–719 nm |
| Analysis Cycle |
1 min demonstrated |
| Operation |
Automated collection–analysis–ablation cycle |
| User Interface | Integrated touchscreen |
| Dimensions | 28 × 28 × 19 cm (H × W × D) |
| Weight | TBD |
Specifications subject to change without notice. Technology is licensed from NIOSH/CDC (Zervaki and Kulkarni,
Patent Application #US20250130140A1)